The Micro Motion™ FVM Fork Viscosity Meter delivers continuous, reliable multivariable measurement of viscosity, density, and temperature through a rugged direct‑insertion design suitable for pipelines, bypass loops, and tanks.
Its hazardous‑area approved, head‑mounted transmitter enables local configuration and display, while advanced internal diagnostics simplify rapid verification of meter health and status.
Micro Motion™ FVM Fork Viscosity Meter

Specs

Density accuracy
±1,0 kg/m3(±0,001 g/cm3)
Meter approvals, calibrations, and certifications
Agencies & types of approval available:
ATEX
CSA
IECEx
Operating density range
0-3000 kg/m3(0-3 g/cm3)
Operating temperature range – long stem
–40 °C to +150 °C (–40 °F to +302 °F)
Operating temperature range – short stem
–50 °C to +200 °C (–58 °F to +392 °F)
Operating viscosity range
0.5 to 20,000 cP
Transmitter output channels
Available output channels (depending on the Electronics Interface option):
One or two externally powered (passive) 4–20 mA outputs
Channel A is 4–20 mA + HART
Channel B is 4–20 mA
One externally powered (passive) Time Period Signal (TPS)output
One externally powered (passive) Discrete output
One Modbus/RS-485 output Can also be ordered with a processor for use with a remote-mount 2700 FOUNDATION fieldbus transmitter
Viscosity accuracy
±0.2 cP (0.5 –10 cP range) then ±1% full scale of the following ranges:
0.5 to 10 cP
10 to 100 cP
100 to 1000 cP

Features

  • Receive reliable and continuous multivariable measurement of viscosity, density and temperature
  • Easily insert directly into pipelines, bypass loops and tanks
  • Expand usage with a hazardous area-approved, head-mounted transmitter that supports local configuration and display
  • Quickly verify meter health and status using advanced internal diagnostics
  • Support multiple simultaneous protocols for connection to DCS, PLC and flow computers