Tank Hub Versions
                                
                                
                                    Single tank version: supports one Rosemount 5900S 2-in-1 gauge or two radar level gauges, type Rosemount 5300, 5408, or 5900 installed on the same tank.
 Multiple tank version: for a Rosemount 5300/5408/5900 system configuration:
- The software supports 16 field devices and 10 tanks per hub.
 - Maximum five Rosemount 5300 or ten 5408 per hub.
 
                                 
                             
                            
                                
                                    Custody Transfer Approval
                                
                                
                                    OIML R85:2008, and national certifications such as PTB, NMi etc.
                                
                             
                            
                                
                                    Safety
                                
                                
                                    • SIL 2 and SIL 3 certified.
• TÜV tested and WHG approved for overfill prevention.
                                
                             
                            
                                
                                    Tankbus
                                
                                
                                    Foundation™ Fieldbus FISCO (self-configuring ).
                                
                             
                            
                                
                                    Field Bus
                                
                                
                                    Rosemount 2410 communicates with a host or a field communication unit via TRL2 Modbus, RS485, 4-20 mA/HART, Enraf, L&J Tankway, Varec Mark/Space, Whessoe, GPE, Sakura, Tokyo Keiso and WirelessHART® protocol.
                                
                             
                            
                                
                                    Relay Outputs
                                
                                
                                    Two SIL certified relays, which can be configured for different tanks and process variables.
                                
                             
                            
                                
                                    Analog Input
                                
                                
                                    One input channel, current range 0-23 mA.
                                
                             
                            
                                
                                    Analog Output
                                
                                
                                    One output channel (SIL certified ), current range 3.5-23 mA (HART).
                                
                             
                            
                                
                                    Emulation Capability
                                
                                
                                    Possible to integrate with legacy tank gauging systems. Supports emulation of other vendors’ communication protocols. See “Field Bus” above.
                                
                             
                            
                                
                                    Display/Interface
                                
                                
                                    Integral digital read-out display: level, level rate, ullage, signal strength, volume (TOV), liquid average temperature, 1-16 spot temperature, vapor average temperature, ambient temperature, free water level, liquid pressure, vapor and air pressure, observed and reference density, and flow rate.
                                
                             
                            
                                
                                    Power Supply
                                
                                
                                    • 24-48 VDC (-15% to +10%).
• 48-240 VAC (-15% to +10%), 50/60 Hz.
                                
                             
                            
                                
                                    Ambient Temperature
                                
                                
                                    -40 to +70 °C (-40 to +158 °F). Minimum start-up temperature is -50 °C (-58 °F).
With LCD display: -25 to 70 °C (-13 to 158 °F).